Proline Promag W 400 EtherNet/IP Electrical connection
Endress+Hauser 43
6. Firmly tighten the cable glands.
7. WARNING! Housing degree of protection may be voided due to insufficient sealing of
the housing. Screw in the screw without using any lubricant. The threads on the
cover are coated with a dry lubricant.
Reverse the procedure to reassemble the sensor.
7.2.2 Connecting the transmitter
L
WARNING
Housing degree of protection may be voided due to insufficient sealing of the housing.
‣
Screw in the screw without using any lubricant. The threads on the cover are coated
with a dry lubricant.
Tightening torques for plastic housing
Housing cover fixing screw 1.3 Nm
Cable entry 4.5 to 5 Nm
Ground terminal 2.5 Nm
PH 2
4 x
1.
2.
3.
4.
5.
10 (0.4)
1
2
L+/L
L-/N
6.
A0021356
15 Connecting the supply voltage and EtherNet/IP
1. Loosen the 4 fixing screws on the housing cover.
2. Open the housing cover.
3. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
4. Strip the cable and cable ends. In the case of stranded cables, also fit ferrules.
5. Connect the cable in accordance with the terminal assignment → 38. For supply
voltage: open the shock protection cover.
6. Firmly tighten the cable glands.
7. WARNING! Housing degree of protection may be voided due to insufficient sealing of
the housing. Screw in the screw without using any lubricant.
Reverse the removal procedure to reassemble the transmitter.