Commissioning Proline Promag W 400 EtherNet/IP
82 Endress+Hauser
Parameter Description Selection Factory setting
Mass unit Select mass unit.
Result
The selected unit is taken from: Mass flow
unit parameter
Unit choose list Country-specific:
• kg
• lb
Density unit Select density unit.
Result
The selected unit applies for:
• Output
• Simulation process variable
• Density adjustment (in Expert menu)
Unit choose list Country-specific:
• kg/l
• lb/ft³
10.6.3 Configuring the communication interface
The Communication submenu guides you systematically through all the parameters that
have to be configured for selecting and setting the communication interface.
Navigation
"Setup" menu → Communication
‣
Communication
MAC address
Default network settings
DHCP client
IP address
Subnet mask
Default gateway
Parameter overview with brief description
Parameter Description User interface / Selection /
User entry
Factory setting
MAC address Displays the MAC address of the measuring
device.
MAC = Media Access Control
Unique 12-digit character
string comprising letters and
numbers, e.g.:
00:07:05:10:01:5F
Each measuring device is given
an individual address.
Default network settings Select whether to restore network settings. • Off
• On
Off
DHCP client Select to activate/deactivate DHCP client
functionality.
Result
If the web server's DHCP client functionality
is activated, the IP address, subnet mask and
default gateway are automatically set.
Identification is via the MAC address
of the measuring device.
• Off
• On
On