Diagnostics and troubleshooting Proline Promag W 400 EtherNet/IP
120 Endress+Hauser
12.12.1 Function scope of "Device reset" parameter
Options Description
Cancel No action is executed and the user exits the parameter.
To delivery settings Every parameter for which a customer-specific default setting was ordered is reset
to this customer-specific value. All other parameters are reset to the factory
setting.
Restart device The restart resets every parameter whose data are in the volatile memory (RAM)
to the factory setting (e.g. measured value data). The device configuration remains
unchanged.
History reset Every parameter is reset to its factory setting.
12.13 Device information
The Device information submenu contains all the parameters that display different
information for identifying the device.
Navigation
"Diagnostics" menu → Device information
‣
Device information
Device tag
Serial number
Firmware version
Device name
Order code
Extended order code 1
Extended order code 2
Extended order code 3
ENP version
IP address
Subnet mask
Default gateway