Technical data Proline Promag W 400 EtherNet/IP
156 Endress+Hauser
Process connections EN 1092-1 (DIN 2501)
3)
DN ≤ 300: fixed flange (PN 10/16/25/40) = form A
DN ≥ 350: fixed flange (PN 6/10/16/25) = flat face
DN 450 to 2000
4)
: fixed flange (PN 6/10/16) = flat face
ASME B16.5
DN 25 to 600 (1 to 24"): fixed flange (Class 150)
DN 350 to 2000 (14 to 78")
4)
: fixed flange (Class 150)
DN 25 to 150 (1 to 6"): fixed flange (Class 300)
AWWA C207
DN 48 to 72": fixed flange (Class D)
DN 48 to 78"
4)
: fixed flange (Class D)
AS 2129
DN 50 to 1200: fixed flange (Table E)
DN 350 to 1200
4)
: fixed flange (Table E)
AS 4087
DN 50 to 1200): fixed flange (PN 16)
DN 350 to 1200
4)
: fixed flange (PN 16)
JIS B2220
DN 50 to 300: fixed flange (10K)
DN 25 to 300: fixed flange (20K)
For information on the materials of the process connections →  154
Surface roughness Electrodes with 1.4435 (316L); Alloy C22, 2.4602 (UNS N06022); tantalum:
≤ 0.3 to 0.5 µm (11.8 to 19.7 µin)
(All data relate to parts in contact with fluid)
16.11 Operability
Local operation Via display module
A0020538
Display elements
4-line display
White background lighting; switches to red in event of device errors
Format for displaying measured variables and status variables can be individually
configured
Permitted ambient temperature for the display: –20 to +50 °C (–4 to +122 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
3) Dimensions as per DIN 2501, DN 65 (2 ½") PN 16 and DN 600 (24") PN 16 only as per EN 1092-1