Operation options Proline Promag W 400 EtherNet/IP
64 Endress+Hauser
Computer settings
User rights User rights are required for TCP/IP and proxy server settings (for changes to
the IP address, subnet mask etc.).
Proxy server settings of the
Web browser
The Web browser setting Use proxy server for LAN must be disabled.
JavaScript JavaScript must be enabled.
If JavaScript cannot be enabled:
enter http://192.168.1.212/basic.html in the address line of the Web
browser. A fully functional but simplified version of the operating menu
structure starts in the Web browser.
When installing a new firmware version: To enable correct data display,
clear the temporary memory (cache) of the Web browser under Internet
options.
Measuring device
Web server Web server must be enabled; factory setting: ON
For information on enabling the Web server →  67
IP address If the IP address of the device is not known:
The IP address can be read out via local operation:
"Diagnostics" menu → Device information → IP address
Communication with the Web server can be established via the standard IP
address 192.168.1.212.
The DHCP function is enabled in the device at the factory, i.e. the device
expects an IP address to be assigned by the network. This function can be
disabled and the device can be set to the standard IP address
192.168.1.212: set top DIP switch No. 2 from OFF → ON.
1
2
3
4
1
2
3
4
1
2
ON OFF
Default Ethernet network settings
IP 192.168.1.212
Write protection
128
64
32
16
8
4
2
1
IP Address
setting
(last octet)
A0023353
Once the DIP switch has been activated, the device must be restarted
before the device uses the standard IP address.
If the standard IP address (top DIP switch No. 2 = ON) is used, there is
no connection to the EtherNet/IP network.
8.4.3 Establishing a connection
Configuring the Internet protocol of the computer
IP address 192.168.1.XXX; for XXX all numerical values except: 0, 212 and 255 → e.g.
192.168.1.213
Subnet mask 255.255.255.0
Default gateway 192.168.1.212 or leave cells empty
1. Switch on the measuring device and connect to the computer via the cable
→  68.