Parameter Prerequsite Description Selection / User
entry
Factory setting
Display language – Set display language. • English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык
(Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• ةّيبَرَعْلا (Arabic)
• Bahasa Indonesia
• ภาษาไทย (Thai)
• tiếng Việt
(Vietnamese)
• čeština (Czech)
English
(alternatively, the
ordered language is
preset in the device)
Display interval – Set time measured values are
shown on display if display
alternates between values.
1 to 10 s 5 s
Display damping – Set display reaction time to
fluctuations in the measured
value.
0.0 to 999.9 s 0.0 s
Header – Select header contents on local
display.
• Device tag
• Free text
Device tag
Header text – Enter display header text. ------------
Separator – Select decimal separator for
displaying numerical values.
• .
• ,
.
Backlight – Switch the local display
backlight on and off.
Only for device version
with onsite display SD03
(touch control)
• Disable
• Enable
Enable
10.7.4 Performing electrode cleaning
The Electrode cleaning circuit wizard guides you systematically through all the
parameters that have to be set for configuring electrode cleaning.
The wizard only appears if the device was ordered with an electrode cleaning circuit.
Navigation
"Setup" menu → Advanced setup → Electrode cleaning circuit
‣
Electrode cleaning circuit
Electrode cleaning circuit
ECC duration
ECC recovery time