Maintenance
Eppendorf Xplorer
®
/Eppendorf Xplorer
®
plus
English (EN)
77
7.5.3.4 Loosening and opening the 1 200 μL multi-channel
The method of loosening the 1 200 µL lower part matches that of loosening the 5 mL or
10 mL lower parts.
1. The piston must be in its basic position before loosening the lower part. Press the
rocker down to move the piston into its basic position.
2. With the ejector pressed down, turn the lower part approximately 30° and pull down to
cancel the magnetic coupling. This releases the lower part from the upper part.
3. View the detached lower part from above. For this, slide the housing case down. Note
the exact position of the red marked parts. The red marked stops on the housing case
and the red marked teeth on the inner part are adjacent. If the stops and teeth are on
top of each other after assembly, the housing case has been attached incorrectly.