EPSON Stylus Photo R1900/R2880/R2000/R2000s/SC-P400 Series  Revision I
Adjustment  Adjustment  147
8. Press the [MEASURE] button. (“----” is displayed on the LCD screen.)
9. Put the tip of the tweezers on the Drive Belt, and flip it downward in that position. 
The “----” displayed on the LCD will become wave pattern during the 
measurement. When it has finished, the measurement result will be displayed by 
“N” (Newton) after the beep. This jig can pick up and measure sounds accurately, 
regardless of the flipping force.
10. Repeating 8 and 9, delicately shift the variable part of the PF Motor mounting 
position to adjust the tension until the tension falls within the allowable standard 
value.
5.2.2  PG Adjustment
When any of the following parts has been removed or replaced, this adjustment must 
be performed to secure the specified clearance between the print surface of the Print 
Head and paper.
 Print Head
 Carriage Unit
 Carriage Shaft
 Parallelism Adjust Bushing (Including the case when just moved it)
In this adjustment, use the same Adjustment Gauge on the left and right sides.
Figure 5-3. Adjustment Gauge
As the Drive Belt is flipped with the tip of tweezers in the following 
steps, carefully choose the flipping position so that the Belt will not 
make contact with the Microphone by reaction of flipping.
A D J U S T M E N T
R E Q U I R E D
Standard Value: 10.5 ± 2N (8.5 ~ 12.5N)
 Even if the Timing Belt is flipped, the LCD screen may not 
change at all. In this case, flip the Timing Belt again after a few 
seconds have passed.
 If measurement results differ greatly from each other, acoustic 
sounds may not be picked up properly in any of the 
measurements. Therefore, flip the Timing Belt again with the 
tweezers, and record the value at which two measurement 
results are approximate. Displaying errors in the range 1/100 to 
5/100, the Measuring Tool has high reliability.
 Some pictures used in this section are Stylus Photo R1800. The 
adjustment method for Stylus Photo R1900/Stylus Photo R2880 
is the same as the one for Stylus Photo R1800.
 To change the PG position, turn the cam each on the left and 
right side of the carriage shaft simultaneously.
 Do not touch the Adjustment Gauge Plate surface with bare 
hands.
 If the Adjustment Gauge Plate surface is stained by ink or, etc. 
wipe it with a soft cloth.
Continuity 
Measuring Point
Adjustment Gauge 
Plate (Top face)