20.4.3.2 Accuracy Values And Ranges
Accuracy Experienced
coefficient
Against Reference
(Cell-Dyn 3700)
Range
WBC r = 1.00 0.5*10
3
/ µl WBC 56.3*10
3
l
RBC r = 0.99 2.2 *10
6
/ µl RBC 6.3 * 10
6
l
HGB r = 0.99 68.8 g/l HGB 173 g/l
HCT r = 0.99 19.5% HCT% 52.6%
PLT r = 0.98 12.5 *10
3
/ µl PLT 699*10
3
l
NEU% r = 0.99 5.8% NEU% 94.6%
LYM% r = 0.99 2.5% LYM% 79.5%
MON% r = 0.90 1.1% MON% 28.1%
EOS% r = 0.95 0.0% EOS% 23.6%
BAS% r = 0.55 0.1% BAS% 4.2%
Note 1: the 5part WBC is not reportable if the total WBC count is below 1 *10
3
/ µl. (The error
grows to high.)