EasyManua.ls Logo

ERBA ELite 5 - Page 150

ERBA ELite 5
165 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ELite 5, User Manual
Page 150 of 159
20.4.3.3 Linearity Values and Ranges
Linearity Coefficient of determination Range
WBC r^2 0.95 1*10
3
/ µl ≤ WBC ≤ 100*10
3
/µl
RBC r^2 0.95 0.4 *10
6
/ µl ≤ RBC ≤ 7.5 * 10
6
/µl
HGB r^2 0.95 13 g/l ≤ HGB ≤ 227 g/l
PLT r^2 0.95 10*10
3
/ µl ≤ PLT ≤ 2000*10
3
/µl
20.4.3.4 Carryover Values and Ranges
Carryover High-to-low effect Range
WBC 0.5% or less LO ≤ 10*10
3
/µl
HI 75*10
3
/µl
RBC 0.5% or less LO ≤ 2.3* 10
6
/µl
HI 7.5* 10
6
/µl
HGB 0.5% or less LO ≤ 90 g/l
HI 270 g/l
PLT 0.5% or less LO ≤ 140 *10
3
/ µl
HI 730*10
3
/ µl

Table of Contents

Related product manuals