20.4.3.3 Linearity Values and Ranges
Linearity Coefficient of determination Range
WBC r^2 0.95 1*10
3
/ µl WBC 100*10
3
l
RBC r^2 0.95 0.4 *10
6
/ µl RBC 7.5 * 10
6
l
HGB r^2 0.95 13 g/l HGB 227 g/l
PLT r^2 0.95 10*10
3
/ µl PLT 2000*10
3
l
20.4.3.4 Carryover Values and Ranges
Carryover High-to-low effect Range
WBC 0.5% or less LO 10*10
3
l
HI 75*10
3
l
RBC 0.5% or less LO 2.3* 10
6
l
HI 7.5* 10
6
l
HGB 0.5% or less LO 90 g/l
HI 270 g/l
PLT 0.5% or less LO 140 *10
3
/ µl
HI 730*10
3
/ µl