Malfunctions – What should you do?
82 MMS
®
Inspection DPM High
Gage always switch-
es off
Small switch-off time
periods (1 min) are set in
the Energy Savings func-
tion.
Increase the switch-
off time periods for
the parameters
Switch off and Switch
off (USB): Main menu
( ) > Gage Settings
> OK > Energy
Savings > OK >
change switch-off time
> OK, see page 71.
Gage no longer
switches off auto-
matically
The Energy Savings
function is disabled
(switch-off time 0 = no
automatic gage switch-
off)
Set the time period for
the automatic gage
switch-off: Main menu
( ) > Gage Settings
> OK > Energy
Savings > OK >
change switch-off time
> OK, see page 71.
No visual alarm sig-
nal at limit violation
The function Optical
Indic. is switched off
Switch on the func-
tion: Main menu ( )
> Gage Settings > OK
> Indication Settings
> Optical Indic. > OK.
The alarm function of the
climatic variable is deac-
tivated.
Activate the alarm
function: Main menu
( ) > Batch Modify >
Alarm Settings >
select climatic variable
> OK > select threshold
> OK set limit value >
OK.
Batch can not delet-
ed
The open batch is
selected for deleting.
Open another batch
first, then you can
delete the desired
batch.