EasyManua.ls Logo

Flexim PIOX S72 Series - Page 120

Flexim PIOX S72 Series
225 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
10 Measurement
10.5 Field calibration PIOX S72*
2022-05-15, UMPIOX_S72xV1-9EN
120
Stop measurement
Press during the measurement.
The following display appears:
•Select Stop measurement.
Press ENTER.
The measurement is stopped.
Input of the reference value
Select the program branch Options.
Press ENTER.
Select a channel for which a field calibration is to be
carried out (here: Channel A).
Press ENTER.
This display will not be indicated if the transmitter has only
one measuring channel.
Select the list item Field calibration c.
Press ENTER.
Select the list item Edit samples.
The sampling number and the time of sampling are displayed.
Press ENTER.
Enter the reference value of the sampling.
Press ENTER.
Fig. 10.4: Scroll list in the program branch Measurement
Options
Channel x
Field calibration
c
Auto calibration Auto activation Edit samples Clear samples
Off Default On
see annex A
Options\Channel A
Options\Channel A\Field calibration c
Notice!
The reference quantity has to be identical with the physical quantity of the fluid selected in the program branch
Parameters.

Table of Contents

Related product manuals