20 Advanced functions
PIOX S72* 20.3 Diagnosis with the help of the snap function
179
UMPIOX_S72xV1-9EN, 2022-05-15
20.3 Diagnosis with the help of the snap function
20.3.1 Configuration
By means of the snap function it is possible to store measuring parameters which are useful for the evaluation of
measuring results or for diagnostic purposes. In order to use of the snap function it has to be configured.
• Select the menu item Configuration.
• Press ENTER.
•Select On to activate the snap function.
• Press ENTER.
•Select Yes to activate the snap ringbuffer.
If the snap ringbuffer is activated, after taking the 51th, snap the oldest snaps are overwritten. If the ringbuffer is
deactivated, up to 50 snaps can be stored.
• Press ENTER.
•Select Yes if the auto snap has to be activated.
If the auto snap is activated, snaps are automatically stored during a measurement failure.
• Press ENTER.
•Select Yes if an event that releases a snap has been parameterized for the event trigger R1.
• Press ENTER.
20.3.2 Taking a snap
• Press and hold C during the measurement until the menu item Execute command is displayed.
• Select the list item Take a snap.
• Press ENTER.
A snap will be taken.
Special functions\Snap\Configuration
Special functions\Snap\Configuration\Snap
Special functions\Snap\Configuration\Snap ringbuffer
Special functions\Snap\Configuration\Auto snap
Special functions\Snap\Configuration\Snap on R1
Measurement\Execute command\Take a snap