11 Troubleshooting
PIOX S72* 11.2 Measuring point selection
131
UMPIOX_S72xV1-9EN, 2022-05-15
11.2 Measuring point selection
Make sure that the recommended min. distance to any disturbance is observed.
Avoid measuring points with deposit formation in the pipe.
Avoid measuring points in the vicinity of deformations and defects on the pipe as well as welds.
Make sure the pipe surface at the selected measuring point is even.
Measure the temperature at the measuring point and make sure that the transducers are suitable for this temperature.
Make sure that the outer pipe diameter is within the measuring range of the transducers.
When measuring on a horizontal pipe, the transducers have to be mounted laterally on the pipe.
A vertical pipe always has to be filled at the measuring point and the fluid should flow upward.
No gas bubbles should form (even bubble-free fluids can form gas bubbles when the fluid expands, e.g., upstream of
pumps and downstream of great cross-section enlargements).
11.3 Maximum acoustic contact
see section 7.2
11.4 Application-specific problems
The measuring points of the physical quantities for analysis are outside the valid range
Check the entered value, see section 10.2.4.
The entered pipe roughness is not appropriate
Check the entered value. Check the entered value considering the pipe condition.
Measurements on pipes made of porous materials (e.g., concrete or cast iron) are only conditionally possible
Contact FLEXIM.
The pipe lining may cause problems during the measurement if it is not firmly attached to the inner pipe wall or
consists of an acoustically absorbing material
Try to measure on a section of the pipe free from lining.
Highly viscous fluids strongly attenuate the ultrasonic signal
The measurement of fluids with a viscosity of > 1000 mm
2
/s is only conditionally possible.
Gas bubbles or solids present in high concentration in the fluid scatter and absorb the ultrasonic signal and thus
attenuate the measuring signal
A measurement is impossible if the value is 10 %. If the proportion is high, but < 10 %, a measurement is only
conditionally possible.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Flexim PIOX S72 Series and is the answer not in the manual?

Flexim PIOX S72 Series Specifications

General IconGeneral
BrandFlexim
ModelPIOX S72 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals