5-16
Table 5-6. STATus Command Summary
Command Description
STATus Path to the Status commands.
:OPERational Path to the Operational Data registers.
:EVENt? Returns value in Operational Data Event register.
:ENABle <DPND> Sets the bits of the Operational Data Enable register.
:ENABle? Returns the vaule in the Operational Data Enable register.
:CONDition? Returns the value in the Operational Condition register.
:QUEStionable Path to the Questionable Data Event register.
:EVENt? Returns value in Questionable Data Event register.
:ENABle <DPND> Sets the bits of the Questionable Data Enable register.
:ENABLe? Returns the vaule in the Questionable Data Enable register.
:CONDition? Returns the value in the Questionable Condition register.
:PRESet Clears all bits in Operation Data Enable and Questionable
Data Enable register.
Table 5-7. Common Command Summary
Command Description
*IDN? Returns manufacturer, model number, serial number and
firmware revision number.
*OPC Sets the OPC bit in the Event Status Register.
*OPC? Returns “1” when all pending operations are complete.
*WAI Prevents the Calibrator from executing command or queries
until all previous remote commands have been executed.
*RST Resets the Calibrator to its initial status.
*TST? Starts a Calibrator self test and returns “0” for pass and “1”
for fail.
*STB? Returns value in the Status Byte Register.
*SRE <value> Sets the Service Request Enable register.
*SRE? Returns value in the Service Request Enable register.
*ESR? Returns value in Event Status register.
*ESE <value> Sets the Event Status register.
*ESE? Returns value in the Event Status Enable register.
*CLS Clears the Event Status register.