Remote Operation
Supported SCPI Commands 5
5-15
Table 5-4. SOURce Command Summary (cont.)
:FUNCtion? Returns the Multimeter Function setting.
:PROBe {OFF|10KV|40KV} Sets HV probe for HIPOT leakage current.
:PROBe? Returns the HV probe for HIPOT leakage current.
:CURRent? Returns the measured current by the ammeter.
:VOLTage? Returns the measured voltage value.
:HIPT Path to HIPOT timer function.
:FUNCtion {DC|AC} Sets the HIPOT timer mode: ac or dc.
:FUNCtion? Returns the Multimeter Function setting.
:TIME? Returns the time interval.
:VOLTage? Returns the measured voltage value.
[:VOLTage]:MAXimum? Returns the maximum measured peak voltage.
:CLEar? Clears the measured time and peak voltage.
Table 5-5. SYSTem Command Summary
Command Description
SYSTem Path to System commands.
:DATE <DNPD>, <DNPD, <DNPD> Sets the system date of the Calibrator.
:DATE? Returns the Date set in the Calibrator.
:TIME <DNPD>, <DNPD, <DNPD> Sets the system Time in the Calibrator.
:TIME? Returns the Time set in the Calibrator.
:ERRor? Returns the first error in the error queue.
:REMote Places the Calibrator in the Remote mode.
:RWLock Places the Calibrator in the Remote mode with lockout.
:LOCal Sets the Calibration to the Local mode.