Remote Operation
Supported SCPI Commands 5
5-19
[SOUR]:SAF:GBOP
Description: This command switches the Calibrator to the Ground Bound
Resistance function and sets the Open mode. If not already selected, it
also switches the Calibrator to the Ground Bound Resistance function.
Example: SAF:GBOP
[SOUR]:SAF:GBOP:VOLT?
Description: This command returns the measured voltage on the open output
terminals. If not already selected, it also switches the Calibrator to the
Ground Bound Resistance function in the Open mode.
Query: SAF:GBOP:VOLT?
[SOUR]:SAF:GBOP[:VOLT]:LIM?
Description: This command returns the maximum allowed voltage on the open
output terminals. If not already selected, it also switches the Calibrator
to the Ground Bound Resistance function in the Open mode.
Query: SAF:GBOP:LIM?
[SOUR]:SAF:GBOP[:VOLT]:MAX?
Description: This command returns the maximum measured peak voltage on the
open output terminals. If not already selected, it also switches the
Calibrator to the Ground Bound Resistance function in the Open
mode.
This value can be cleared using the SAF:GBOP:CLEar command.
Query: SAF:GBOP:MAX?
[SOUR]:SAF:GBOP[:VOLT]:CLE
Description: This command clears the measured peak voltage on the open output
terminals. If not already selected, it also switches the Calibrator to the
Ground Bound Resistance function.
Example: SAF:GBOP:CLE
[SOUR]:SAF:HRES[:LEV](?) [<DNPD>]
Description: This command sets the resistance value for the High Resistance
Source function. If not already selected, it also switches the Calibrator
to the High Resistance Source function.
Parameters: <DNPD> The value of resistance expressed in ohms.
Table 4-2 in Chapter 4 lists the acceptable
range of resistance values.
Example: SAF:HRES 1.052E6 High Resistance Source function with
1.052 MΩ.
Query: SAF:HRES? Returns the value of the selected resistance.