5320A
Users Manual
5-20
[SOUR]:SAF:HRES:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the High Resistance LO
terminal to or from ground. If not already selected, it also switches the
Calibrator to the High Resistance Source function.
Parameters: <CPD> FL = floating output
GRO = grounded output
Example: SAF:HREs:LOW GRO Grounds the output LO terminal.
Query: SAF:HRES:LOW? Returns the grounded state of the Ouput LO
terminal, GRO or FLO.
[SOUR]:SAF:HRES:MULT(?) <CPD> {ON | OFF}
Description: This command enables the probe (high resistance multiplier). If
enabled, the selected resistance value is multiplied by 1000. If not
already selected, it also switches the Calibrator to the High Resistance
Source function.
Parameters: <CPD> ON = enables the high resistance multiplier.
OFF = disables the high resistance
multiplier.
Example: SAF:HRES:MULT ON Enable high resistance multiplier.
Query: SAF:HRES:MULT? Returns the state of the high resistance
multiplier. ON = enabled, OFF = disabled.
[SOUR]:SAF:HRES:RINP(?) <DNPD>
Description: This command sets the input resistance of the Sense terminal. If not
already selected, it also switches the Calibrator to the High Resistance
Source function.
Parameters: <DNPD> The value if resistance expressed in ohms.
Default value is 0 and range is limited to
100 MΩ.
Example: SAF:HRES:RINP 10.52E6 Sets serial resistance to 10.52 MΩ.
Query: SAF:HRES:RINP? Returns the value of the serial resistance.
[SOUR]:SAF:HRES:VOLT?
Description: This command returns the measured voltage on the resistance. If not
already selected, it also switches the Calibrator to the High Resistance
Source function.
Query: SAF:HRES:VOLT?
[SOUR]:SAF:HRES[:VOLT]:LIM?
Description: This command returns maximum allowed value of voltage on the
resistance. If not already selected, it also switches the Calibrator to the
High Resistance Source function.
Query: SAF:HRES:LIM?