Remote Operation
Supported SCPI Commands 5
5-21
[SOUR]:SAF:HRES[:VOLT]:MAX?
Description: This command returns the maximum measured peak voltage on the
resistance. If not already selected, it also switches the Calibrator to the
High Resistance Source function.
This value can be cleared using the SAF:HRES:CLEar command.
Query: SAF:HRES:MAX?
[SOUR]:SAF:HRES[:VOLT]:CLE
Description: This command clears the measured peak voltage on the resistance. If
not already selected, it also switches the Calibrator to the High
Resistance Source function.
Example: SAF:HRES:CLE
[SOUR]:SAF:HRF(?)
Description: This command selects the high resistance source function with the
100 GΩ fixed resistance.
Example: SAF:HRF Selects 100 GΩ fixed resistance.
Query: SAF:HRF? Returns the value of the fixed resistance.
[SOUR]:SAF:HRF:LOW(?) {FLO | GRO}
Description: This command connects or disconnects the High Resistance LO
terminal to or from GND terminal. If not already selected, it also
switches the Calibrator to the High Resistance Fixed 100 GΩ
resistance source function.
Parameters: <CPD> FL = floating output
GRO = grounded output
Example: SAF:HRF:LOW GRO Grounds the High Resistance LO terminal.
Query: SAF:HRF:LOW? Returns the grounded state of the High
Resistance LO terminal, GRO or FLO.
[SOUR]:SAF:HRF:VOLT?
Description: This command returns the measured voltage across the selected fixed
reistance. If not already selected, it also switches the Calibrator to the
High Resistance Fixed 100 GΩ resistance source function.
Query: SAF:HRF:VOLT? Returns the measured voltage across the
fixed resistance.
[SOUR]:SAF:HRF:LIM?
Description: This command returns the maximum allowed voltage across the
selected fixed reistance. If not already selected, it also switches the
Calibrator to the High Resistance Fixed 100 GΩ resistance source
function.
Query: SAF:HRF:LIM? Returns the maximum allowed voltage
across the fixed resistance.