Remote Operation
Supported SCPI Commands 5
5-23
This value can be cleared using the SAF:HRSH:CLEar command.
Query: SAF:HRSH:MAX? Returns the maximum current value.
[SOUR]:SAF:HRSH[:CURR]:CLE
Description: This command clears the measured peak current flowing through the
shorted terminals. If not already selected, it also switches the
Calibrator to the High Resistance Fixed 100 GΩ resistance source
function in the Short mode.
Example: SAF:HRSH:CLE
[SOUR]:SAF:LRES[:LEV](?) [<DNPD>]
Description: This command switches the calibrator to Low Resistance Source
function and sets the resistance value to the value contained in DNPD.
Parameters: <DNPD> The value of the resistance expressed in ohms.
Table 4-1 in Chapter 4 lists the acceptable range
of values.
Example: SAF:LRES 105.2 Low Resistance Source function with 105.2 Ω.
Query: SAF:LRES? Returns the value of the selected resistance.
[SOUR]:SAF:LRES:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the Low Resistance LO
terminal to or from GND terminal. If not already selected, it also
switches the Calibrator to the Low Resistance Source function.
Parameters: <CPD> FL = floating output
GRO = grounded output
Example: SAF:LRES:LOW GRO Grounds the Low Resistance LO terminal.
Query: SAF:LRES:LOW? Returns the grounded output state, GRO or
FLO.
[SOUR]:SAF:LRES:SENS(?) <CPD> {ON | OFF}
Description: This command selects 4-wire Low Resistance mode (remote sense). If
not already selected, it also switches the Calibrator to the Low
Resistance Source function.
Parameters: <CPD> ON = 4-wire mode
OFF = 2-wire mode
Example: SAF:LRES:SENS ON Low Resistance Source function with
105.2 Ω.
Query: SAF:LRES:SENS? Returns the state of the 4-wire mode, ON or
OFF.
[SOUR]:SAF:LRES:CURR?
Description: This command returns the measured value of current flowing through
the resistance. If not already selected, it also switches the Calibrator to
the Low Resistance Source function.