Query: SAF:LRES:CURR? Returns the measured current value.
[SOUR]:SAF:LRES[:CURR]:LIM?
Description: This command returns the maximum current allowed to flow through
the resistance. If not already selected, it also switches the Calibrator to
the Low Resistance Source function.
Query: SAF:LRES:LIM? Returns the maximum current value.
[SOUR]:SAF:LRES[:CURR]:MAX?
Description: This command returns the maximum measured peak current flowing
through the selected resistance. If not already selected, it also switches
the Calibrator to the Low Resistance Source function.
This value can be cleared using the SAF:LR:CLEar command.
Query: SAF:LRES:MAX? Returns the maximum current value.
[SOUR]:SAF:LRES[:CURR]:CLE
Description: This command clears the measured peak current flowing through the
selected resistance. If not already selected, it also switches the
Calibrator to the Low Resistance Source function.
Example: SAF:LRES:CLE
[SOUR]:SAF:LROP
Description: This command switches the Calibrator to the Low Resistance Source
function and sets the Open mode.
Example: SAF:LROP
[SOUR]:SAF:LROP:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the Low Resistance LO
terminal to or from GND terminal. If not already selected, it also
switches the Calibrator to the Low Resistance Source function in the
Open mode.
Parameters: <CPD> FL = floating output
GRO = grounded output
Example: SAF:LROP:LOW GRO Grounds the Low Resistance LO terminal.
Query: SAF:LROP:LOW? Returns the grounded output state, GRO or
FLO.
[SOUR]:SAF:LROP:VOLT?
Description: This command returns the measured voltage on the open output
terminals. If not already selected, it also switches the Calibrator to the
Low Resistance function in the Open mode.
Query: SAF:LROP:VOLT?