Remote Operation
Supported SCPI Commands 5
5-25
[SOUR]:SAF:LROP[:VOLT]:LIM?
Description: This command returns the maximum allowed voltage on the open
output terminals. If not already selected, it also switches the Calibrator
to the Low Resistance function in the Open mode.
Query: SAF:LROP:LIM?
[SOUR]:SAF:LROP[:VOLT]:MAX?
Description: This command returns the maximum measured peak voltage on the
open output terminals. If not already selected, it also switches the
Calibrator to the Low Resistance function in the Open mode.
This value can be cleared using the SAF:GBOP:CLEar command.
Query: SAF:LROP:MAX?
[SOUR]:SAF:LROP[VOLT]:CLE
Description: This command clears the measured peak voltage on the open output
terminals. If not already selected, it also switches the Calibrator to the
Low Resistance function in the Open mode.
Example: SAF:LROP:CLE
[SOUR]:SAF:LRSH
Description: This command switches the Calibrator to the Low Resistance Source
function in the Short mode.
Example: SAF:LRSH
[SOUR]:SAF:LRSH:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the Low Resistance LO
terminal to or from GND terminal. If not already selected, it also
switches the Calibrator to the Low Resistance Source function in the
Short mode.
Parameters: <CPD> FL = floating output
GRO = grounded output
Example: SAF:LRSH:LOW GRO Grounds the Low Resistance LO terminal.
Query: SAF:LRSH:LOW? Returns the grounded output state, GRO or
FLO.
[SOUR]:SAF:IDAC
Description: This command switches the Calibrator to the Active Leakage Current
function.
Example: SAF:IDAC Active Leakage Current function selected with
parameters set previously.
[SOUR]:SAF:IDAC[:CURR]:NOM (?) <DNPD>
Description: This command sets the nominal current value to the value specified in
DNPD. If not already selected, it also switches the Calibrator to the
Active Leakage Current function.