5320A
Users Manual
5-26
Parameters: <DNPD> The value of nominal current expressed in amps.
Refer to the Leakage Current specifications in
Chapter 1 for the acceptable range of values for
this function.
Example: SAF:IDAC:NOM 0.01 Active Leakage Current function with a
nominal current of 10 mA.
Query: SAF:IDAC:NOM? Returns the nominal leakage current value.
Note: Nominal leakage current is valid for all leakage current modes (IDA,
IDP, IDS and IDD).
[SOUR]:SAF:IDS
Description: This command switches the Calibrator to the Substitute Leakage
Current function.
Example: SAF:IDS Substitute Current function selected with
previously set parameters.
[SOUR]:SAF:IDS[:CURR]:NOM (?) <DNPD>
Description: This command sets the nominal leakage current to the value specified
in DNPD. If not already selected, it also switches the Calibrator to the
Substitute Leakage Current function.
Parameters: <DNPD> The value of leakage current expressed in
amps. Refer to the Leakage Current
specifications in Chapter 1 for the
acceptable range of values for this function.
Example: SAF:IDS:NOM 0.01 Substitute Leakage Current function with
10 mA nominal current setting.
Query: SAF:IDS:NOM? Returns the value of the nominal current
setting.
Note: Nominal leakage current is valid for all leakage current modes (IDA,
IDP, IDS and IDD).
[SOUR]:SAF:IDS:ROUT(?) <DPND>
Description: This command sets the output UUT resistance to the value specified in
DNPD. If not already selected, it also switches the Calibrator to the
Substitute Leakage Current function.
Parameters: <DNPD> The value of UUT output resistance
expressed in ohms.
Example: SAF:IDS:ROUT 2000 Sets the output resistance of the UUT to
2000 Ω.
Query: SAF:IDS:ROUT? Returns the value of the UUT output
resistance setting.
[SOUR]:SAF:IDP
Description: This command switches the Calibrator to the Passive Leakage Current
function.