Remote Operation
Supported SCPI Commands 5
5-27
Example: SAF:IDP Passive Current function selected with parameters
set previously.
[SOUR]:SAF:IDP[:CURR]:NOM (?) <DNPD>
Description: This command sets the nominal leakage current to the value specified
in DNPD. If not already selected, it also switches the Calibrator to the
Passive Leakage Current function.
Parameters: <DNPD> The value of leakage current expressed in
amps. Refer to the Leakage Current
specifications in Chapter 1 for the
acceptable range of values for this function.
Example: SAF:IDP:NOM 0.01 Passive Leakage Current function with
10 mA nominal current setting.
Query: SAF:IDP:NOM? Returns the value of the nominal current
setting.
Note: Nominal leakage current is valid for all leakage current modes (IDA,
IDP, IDS and IDD).
[SOUR]:SAF:IDP[:CURR]:INST?
Description: This command returns the instantaneous measured current flowing
through the calibrator. If not already selected, it also switches the
Calibrator to the Passive Leakage Current function.
Query: SAF:IDP:INST? Returns the value of the instantaneous
current flowing through the Calibrator.
[SOUR]:SAF:IDP[:CURR]:RES?
Description: This command returns the measured current flowing through the
UUT. If not already selected, it also switches the Calibrator to the
Passive Leakage Current function.
Query: SAF:IDP:RES? Returns the value of the resulting measured
UUT current.
[SOUR]:SAF:IDD
Description: This command switches the Calibrator to the Differential Leakage
Current function.
Example: SAF:IDD Differential Current function selected with
parameters set previously.
[SOUR]:SAF:IDD[:CURR]:NOM (?) <DNPD>
Description: This command sets the nominal leakage current to the value specified
in DNPD. If not already selected, it also switches the Calibrator to the
Differential Leakage Current function.
Parameters: <DNPD> The value of leakage current expressed in
amps. Refer to the Leakage Current
specifications in Chapter 1 for the
acceptable range of values for this function.