Remote Operation
Supported SCPI Commands 5
5-29
Query: SAF:RCDT:POS? Returns the index value of the selected resistance.
Note: The selected resistance is valid for RCD Current function as well.
[SOUR]:SAF:RCDT[:CURR]:NOM(?) <DNPD>
Description: This command sets the nominal trip current value to the value
specified by DNPD. If not already selected, it also switches the
Calibrator to the RCD Time function.
Parameters: <DNPD> The value of trip current expressed in amps.
Refer to the RCD specifications in Chapter 1
for the acceptable range of values for this
function.
Example: SAF:RCDT:NOM 1 RCD Trip Time function with a nominal trip
current of 1 amp.
Query: SAF:RCDT:NOM? Returns the value of the nominal trip current
setting.
Note: The selected nominal current value is valid for the RCD Current
function as well.
[SOUR]:SAF:RCDT[:CURR]:MULT (?) <CPD> {0.5X | 1X | 1.4X | 2X | 5x}
Description: This command sets the multiplier of nominal trip current to the value
specified by CPD. If not already selected, it also switches the
Calibrator to the RCD Time function.
Parameters: <CPD> 0.5X = 0.5 times nominal trip current
1X = 1 times nominal trip current
1.4X = 1.4 times nominal trip current
2X = two times nominal trip current
5X = five times nominal trip current
Example: SAF:RCDT:MULT 2X RCD Trip Time function with a times 2
multiplier of nominal trip current.
Query: SAF:RCDT:MULT? Returns the multiplier value for the nominal
trip current setting.
[SOUR]:SAF:RCDT[:CURR]:LEV (?) <CPD> {5 % | 30 % | 60 % | 75% | 90 % |
100 % | 120 %}
Description: This command sets the percent of nominal trip current level to the
value specified by CPD. If not already selected, it also switches the
Calibrator to the RCD Time function.
Parameters: <CPD> The percentage of trip current.
Example: SAF:RCDT:LEV 60 % RCD Trip Time function with a current level
of 60 %.
Query: SAF:RCDT:LEV? Returns the current level setting for the RCD
trip time function.
[SOUR]:SAF:RCDT:[VOLT]:LIN?
Description: This command returns the power supply voltage. If not already
selected, it also switches the Calibrator to the RCD Time function.