5320A
Users Manual
5-30
Query: SAF:RCDT:LINE? Returns the measured value of the voltage
power supply.
[SOUR]:SAF:RCDT:[VOLT]:TOUC?
Description: This command returns the measured touch voltage. If not already
selected, it also switches the Calibrator to the RCD Time function.
Query: SAF:RCDT:TOUC? Returns the measured value of the touch
voltage.
[SOUR]:SAF:RCDT:POL?
Description: This command returns the type of test signal generated by the UUT
{POS | NEG | SYMP | SYMN | DCP | DCN | NO}. If not already
selected, it also switches the Calibrator to the RCD Time function.
Query: SAF:RCDT:POL? Returns the type of UUT test signal.
Types are:
W Test current is pulse (dc) with positive polarity (POS)
X Test current is pulse (dc) with negative polarity (NEG)
I Test current is symmetrical (ac) with positive phase (SYMP)
J Test current is symmetrical (ac) with negative phase (SYMN)
a Test current is dc with positive polarity (DCP)
b Test current is dc with negative polarity (DCN)
NO is returned when the generated test signal is not valid.
[SOUR]:SAF:RCDC
Description: This command switches the Calibrator to RCD Trip Current function.
Example: SAF:RCDC Set RCD Trip Current function.
[SOUR]:SAF:RCDC:CURR?
Description: This returns the trip current. If not already selected, it also switches
the Calibrator to the RCD Current function.
Query: SAF:RCDC:CURR? Returns the current trip level.
[SOUR]:SAF:RCDC:RPOS(?) <DNPD>
Description: This command selects one of the 16 resistances available in the RCD
Current function. The index of resistance is from 0 (25 mΩ) to 15
(2 kΩ). If not already selected, it also switches the Calibrator to the
RCD Current function.
Parameters: <DNPD> Index number of the desired resistance. The range
is 0 (25 mΩ) to 15 (2 kΩ).
Example: SAF:RCDC:POS 2 RCD Current function with 100 mΩ (position 2).
Query: SAF:RCDC:POS? Returns the index value of the selected resistance.
Note: The selected resistance is valid for the RCD Time function as well.