Remote Operation
Supported SCPI Commands 5
5-31
[SOUR:SAF:RCDC[:CURR]:NOM(?) <DPND>
Description: This command sets the nominal trip current value to the value
specified by DNPD. If not already selected, it also switches the
Calibrator to the RCD Current function.
Parameters: <DNPD> The value of trip current expressed in amps.
Example: SAF:RCDC:NOM 1 RCD Trip Current function with a nominal
trip current of 1 amp.
Query: SAF:RCDC:NOM? Returns the value of the nominal trip current
setting.
Note: The selected nominal current value is valid for the RCD Time function
as well.
[SOUR:SAF:RCDC[:CURR]:MAX?
Description: This command returns the maximum (peak) value of trip current. Use
the SAF:RCDC:CLEar command to clear the maximum value.
Calibrator returns the current value using standard exponential format.
For example, 20 mA is returned as 2.000000e-002.
Example: SAF:RCDC:MAX?
Note: Command switches the calibrator in RCD Trip Current function mode.
[SOUR:SAF:RCDC[:CURR]:CLE
Description: This command clears the maximum (peak) value of trip current
Example: SAF:RCDC:CLE
Note: Command switches the calibrator in RCD Trip Current function mode.
[SOUR]:SAF:RCDC:[VOLT]:LIN?
Description: This command returns the power supply voltage. If not already
selected, it also switches the Calibrator to the RCD Current function.
Query: SAF:RCDT:LIN? Returns the measured value of the voltage
power supply.
[SOUR]:SAF:RCDC:[VOLT]:TOUC?
Description: This command returns the measured touch voltage. If not already
selected, it also switches the Calibrator to the RCD Current function.
Query: SAF:RCDT:TOUC? Returns the measured value of the touch
voltage.
[SOUR]:SAF:RCDC:POL?
Description: This command returns the type of test signal generated by the UUT
{POS | NEG | SYMP | SYMN | DCP | DCN | NO}. If not already
selected, it also switches the Calibrator to the RCD Current function.
Query: SAF:RCDC:POL? Returns the type of UUT test signal.
Types are:
W Test current is pulse (dc) with positive polarity (POS)