Remote Operation
Supported SCPI Commands 5
5-35
[SOUR]:SAF:LOOP[:CURR]:PFC?
Description: Returns the value of prospective fault current. If not already selected,
it also switches the Calibrator to the Loop Impedance function.
Query: SAF:LOOP:PFC?
[SOUR]:SAF:LOOP:MAN(?) <DNPD>
Description: This command sets the manual value of loop residual impedance. If
not already selected, it also switches the Calibrator to the Loop
Impedance function.
Parameters: <DNPD> The value of residual impedance in ohms.
Example: SAF:LOOP:MAN 0.87 Set manual compensation value to 0.87 Ω.
Query: SAF:LOOP:MAN? Returns the manual compensation value.
[SOUR]:SAF:LOOP:SCAN?
Description: Returns the scanned loop residual impedance value. If not already
selected, it also switches the Calibrator to the Loop Impedance
function.
Query: SAF:LOOP:SCAN?
[SOUR]:SAF:VOLT[: LEV](?) [<DNPD>]
Description: This command sets the output voltage to the value specified in DNPD.
If not already selected, it also switches the Calibrator to the Voltage
Calibration function.
Parameters: <DNPD> The value of generated voltage expressed in
volts.
Example: SAF:VOLT 100 Set voltage calibration mode with a value
100 volts.
Query: SAF:VOLT? Returns the set voltage value.
[SOUR]:SAFE:VOLT:CURR?
Description: This command returns the measured load current value. If not already
selected, it also switches the Calibrator to the Voltage Calibration
function.
Query: SAF:VOLT:CURR?
[SOUR]:SAF:VOLT:FREQ(?) <DNPD>
Description: This command sets the frequency of the generated signal for ac
Voltage Calibration function to the value specified in DNPD. If not
already selected, it also switches the Calibrator to the Voltage
Calibration function.
Parameters: <DNPD> The value of generated signal frequency
expressed in Hertz.
Example: SAF:VOLT:FREQ 60 Sets the generated signal frequency to
60 Hz.
Query: SAF:VOLT? Returns the set frequency value.