[SOUR]:SAF:VOLT:FUNC(?) <CPD> {DC | AC}
Description: This command sets the voltage generated by the Calibrator to either ac
or dc Voltage. If not already selected, it also switches the Calibrator to
the Voltage Calibration function.
Parameters: <CPD> Type of voltage generation.
DC = dc voltage.
AC = ac voltage.
Example: SAF:VOLT:FUNC DC Sets voltage calibration function to dc
voltage generation.
Query: SAF:VOLT:FUNC? Returns the type of voltage generation
selected.
[SOUR]:SAF:VOLT:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the Ouput LO terminal to or
from GND terminal. If not already selected, it also switches the
Calibrator to the Voltage Calibration function.
Parameters: <CPD> FLO = Floating output
GRO = Grounded output
Example: SAF:VOLT:LOW GRO Sets the LO output terminal to grounded for
voltage generation.
Query: SAF:VOLT:LOW? Returns the grounded output state, GRO or
FLO.
[SOUR]:SAF:MET:FUNC(?) <CPD> {DC | AC}
Description: This command switches the meter function between ac and dc
measurements. If not already selected, it also switches the Calibrator
to the Meter function.
Parameters: <CPD> Type of voltage measurement.
DC = dc measurements.
AC = ac measurements.
Example: SAF:MET:FUNC DC Sets the Multimeter function to dc
measurements.
Query: SAF:MET:FUNC? Returns the multimeter measurement
selection.
[SOUR]:SAF:MET:PROB(?) <CPD> {OFF | 10KV | 40KV}
Description: This command selects whether or not a high voltage probe is used for
the meter function and if it is, which probe is used. If not already
selected, it also switches the Calibrator to the Meter function.
Parameters: <CPD> Type of voltage measurement.
OFF = without probe.
10KV = multimeter with 10 kV probe.
40KV = multimeter with 40 kV probe.
Example: SAF:MET:PROB 10KV Sets the multimeter to measure with a 10 kV
probe.