Remote Operation
Supported SCPI Commands 5
5-37
Query: SAF:MET:PROB? Returns whether the multimeter is
measuring with a probe and if so, which
probe is being used.
[SOUR]:SAF:MET:LOW(?) <CPD> {FLO | GRO}
Description: This command connects or disconnects the Multimeter COM terminal
to or from GND terminal. If not already selected, it also switches the
Calibrator to the Meter function.
Parameters: <CPD> FLO = floating output
GRO = grounded output
Example: SAF:MET:LOW GRO Sets the LO multimeter terminal to
grounded.
Query: SAF:MET:LOW? Returns the grounded output state of the
COM Multimeter terminal, GRO or FLO.
[SOUR]:SAF:MET:CURR?
Description: Returns the measured value of current. If not already selected, it also
switches the Calibrator to the Meter function.
Query: SAF:MET:CURR?
[SOUR]:SAF:MET:POW?
Description: Returns the measured value of power. If not already selected, it also
switches the Calibrator to the Meter function.
Query: SAF:MET:POW?
[SOUR]:SAF:MET:VOLT?
Description: Returns the measured value of voltage. If not already selected, it also
switches the Calibrator to the Meter function.
Query: SAF:MET:VOLT?
[SOUR]:SAF:HIPL:FUNC(?) <CPD> {DC | AC}
Description: This command selects the dc or ac HIPOT leakage current mode.
Parameters: <CPD> ac = sets ac HIPOT leakage current mode
dc = sets dc HIPOT leakage current mode
Example: SAF:HIPL:FUNC AC Sets the ac HIPOT leakage current mode.
Query: SAF:HIPL:FUNC? Returns ac or dc state of the function.
Note: The command switches the calibrator in HIPOT leakage current
function mode.
[SOUR]:SAF:HIPL:PROB(?) <CPD> {OFF | 10KV | 40KV}
Description: This command selects whether or not a high voltage probe is used for
the meter function and if it is, which probe is used. If not already
selected, it also switches the Calibrator to the Meter function.
Parameters: <CPD> Type of HIPOT leakage current
measurement.