A-1
Appendix A
Errors
Introduction
Table A-1 lists the errors the Calibrator can generate. Each error has an ID, a message
and a decritptive explanation.
Table A-1. List of Errors
ID Message Description
701,702 Output/Input overloaded Input or output signal over specified limits. Decrease
signal level.
703 Temperature too high. Calibrator power stage overheated. Disconnect external
load.
704,705,706 Output/Input overloaded Input or output signal over specified limits. Decrease
signal level.
501 Eeprom write. Eeprom write failed.
502 Eeprom read. Eeprom read failed.
503 Eeprom error. Eeprom data lost.
722 Unexpected crossing. Error in internal communication.
721 Unknown function. Error in internal communication.
731 Calibrator not ready Error in internal communication.
732 Internal cpu RESET Calibrator will be restarted.
742 Interface data Error in internal communication.
744 Measurement not
available
Error in internal communication.
743 Interface receive Error in internal communication.
745 Ifc ready timeout. Error in internal communication.