Remote Operation
Supported SCPI Commands 5
5-9
Table 5-4. SOURce Command Summary
Command Description
[SOURce] Path to SOURCE subsystem.
:SAFety Path to SAFETY modes.
:MODE? Returns mode the Calibrator is set to.
:GBResistance Sets Calibrator to Ground Bond Resistance mode.
[:LEVel] <DNPD> Sets the value for the Ground Bond Resistance mode.
[:LEVel]? Returns the set value for the Ground Bond Resistance
mode.
:POSition <DNPD> Sets the Ground Bond Resistance value to one of the 16
resistors.
:POSition? Returns the selected resistor for the Ground Bond
Resistance mode (0 = 25 mΩ to 15 = 2 kΩ).
:CURRent? Returns the current flowing through the selected resistor.
[:CURR]:MAXimum? Returns the measured peak current flowing through the
selected resistance.
[:CURR]:CLEar? Clears the measured peak current flowing through the
selected resistance.
[:CURR]:LIMit? Returns the maximum current allowed to flow through the
resistance.
:GBOPen Sets the Ground Bond Resistance function to the Open
mode.