5320A
Users Manual
5-10
Table 5-4. SOURce Command Summary (cont.)
Command Description
:VOLTage? Returns the measured voltage on the open output
terminals.
[:VOLTage] Path to the Ground Bond Resistance function’s voltage
parameters.
[:VOLTage]:MAXimum? Returns the maximum measured peak voltage on the
open output terminals.
[:VOLTage]:CLEar? Clears the measured peak voltage on the open output
terminals.
[:VOLTage]:LIMit? Returns the maximum allowed voltage on the open output
terminals.
:HRESistance Sets Calibrator to High Resistance mode.
[:LEVel] <DNPD> Sets the High Resistance value.
[:LEVel]? Returns the set value for the High Resistance mode.
:LOW {GROund|FLOat} Connects or disconnect the Low terminal to ground.
:LOW? Returns status of the Low Terminal grounding.
:MULTiplier {ON|OFF} Turns high resistance multiplier ON or OFF .
:MULTiplier? Returns ON when multiplier is selected and OFF when it is
not.
:RINP <DNPD> Sets the sense terminal input resistance
:RINP? Returns the resistance set at the sense terminal
:VOLTage? Returns the voltage applied to the selected resistance.
[:VOLT]:LIMit? Returns the maximum voltage allowed on the resistance.
:HRFix Sets Calibrator to High Resistance Fixed resistance (100
GΩ) function.
[:LEVel]? Returns the set value for the Resistance.
:LOW {GROund|FLOat} Connects or disconnect the Low terminal to ground.
:LOW? Returns status of the Low Terminal grounding.
[:VOLTage] Path to High Resistance Voltage parameters.
[:VOLTage]? Returns the measured peak voltage on the open output
terminals.
[:VOLTage]:MAXimum? Returns the maximum allowed voltage across the
resistance.
[:VOLTage]:CLEar? Clears the measured peak voltage on the output
terminals.
[:VOLTage]:LIMit? Returns the maximum voltage allowed across the
resistance.