Remote Operation
Supported SCPI Commands 5
5-11
Table 5-4. SOURce Command Summary (cont.)
Command Description
:HRSHort Sets Calibrator to High Resistance function in the Short
mode.
:LOW {GROund|FLOat} Connects or disconnect the Low terminal to ground.
:LOW? Returns status of the Low Terminal grounding.
:CURRent? Returns the current flowing through the selected resistor.
[:CURRent] Path to High Resistance Short mode current parameters.
[:CURR]:MAXimum? Returns the maximum measured peak current flowing
through the shorted terminals.
[:CURR]:CLEar? Clears the maximum measured peak current flowing
through the shorted terminals.
[:CURR]:LIMit? Returns the maximum current allowed to flow through the
resistance.
:LROPen Sets Calibrator to Low Resistance function in the Open
mode.
:LOW {GROund|FLOat} Connects or disconnect the Low terminal to ground.
[:LOW]? Returns status of the Low Terminal grounding.
[:VOLTage] Returns the measured voltage on the open output
terminals.
[:VOLTage]:MAXimum? Returns the maximum measured peak voltage on the
open output terminals.
[:VOLTage]:CLEar? Clears the measured peak voltage on the open output
terminals.
[:VOLTage]:LIMit? Returns the maximum voltage allowed across the
resistance.
:LRSHort Sets Calibrator to Low Resistance in Short mode.
:LOW {GROund|FLOat} Returns status of the Low Terminal grounding.
:LOW? Returns the current flowing through the selected resistor.
:IDACtive Sets Calibrator to Active Leakage Current Mode.
[:CURRent]:NOMinal <DNPD> Sets the Active Leakage Current value.
[:CURRent]:NOMinal? Returns the nominal Active Leakage Current Value.
:IDSubstitute Sets Calibrator to Substitute Leakage Current Mode.
[:CURRent]:NOMinal <DNPD> Sets the Substitute Leakage Current value.
[:CURRent]:NOMinal? Returns the nominal Substitute Leakage Current Value.
[:CURRent]:ROUT Set the Substitute Leakage Current R
OUT
value.
[:CURRent]:ROUT? Returns the Substitute Leakage Current R
OUT
value.
:LRESistance Sets Calibrator to Low Resistance mode.

Table of Contents

Other manuals for Fluke 5320A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 5320A and is the answer not in the manual?

Fluke 5320A Specifications

General IconGeneral
BrandFluke
Model5320A
CategoryTest Equipment
LanguageEnglish

Related product manuals