DTX Series CableAnalyzer
Technical Reference Handbook
5-4
Table 5-1. Coaxial Cable Test Settings (cont.)
Setting Description
SETUP > Instrument
Settings > Store Plot
Data
Standard: The tester displays and saves plot data for insertion loss. The tester saves data for the
frequency range required by the selected test limit.
Extended: The tester saves data beyond the frequency range required by the selected test limit.
No: Plot data is not saved, which lets you save more results. Saved results show only worst margins
and worst values for each pair.
SPECIAL FUNCTIONS >
Set Reference
The tester must be referenced to the smart remote the first time the two units are used together.
You should also set the reference every 30 days. See “Setting the Reference” on page 5-1.
Settings for saving tests
Cable ID Source, Current Folder, Result Storage Location (DTX-1800, DTX-1200), Operator, Site,
and Company. See “Preparing to Save Tests” in on page 2-1.
SETUP > Instrument
Settings > Power Line
Frequency
Set to the power frequency in the area where the tester will be used. This setting helps keep ac
noise (50 Hz or 60 Hz) from affecting resistance measurements.