EasyManua.ls Logo

Fluke DTX-MFM2 - DTX-MFM2;SFM2;GFM2 Fiber Module Specifications

Fluke DTX-MFM2
320 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Specifications
Specifications
13-31
13
DTX-MFM2/SFM2/GFM2 Fiber Module
Specifications
Power Meter Specifications
Input connector Interchangeable connector adapter (SC standard)
Detector type InGaAs
Calibrated wavelengths 850 nm, 1310 nm, 1550 nm
Power measurement range 0 dBm to -60 dBm (1300/1310 nm and 1550 nm)
0 dBm to -52 dBm (850 nm)
Display resolution dB, dBm: 0.01
Linear (μW): >400, >40, >4, >0.4, 0.4: 1, 0.1, 0.01, 0.001, 0.0001
Power measurement uncertainty (accuracy)
± 0.25 dB
1
Measurement linearity
(18 °C to 28 °C constant temperature)
1300/1310 nm and 1550 nm: ± 0.1 dB
2
850 nm: ± 0.2 dB
3
Re-calibration period 1 year
Display update rate 1 reading per second
1. Under the following conditions:
Power level: -20 dBm, continuous wave
At 850 nm: 62.5/125 mm fiber with 0.275 NA
At 1310 nm and 1550 nm: 9/125 mm
Ambient temperature: 23
o
C ±5
o
C
2. Linearity for 1310 nm and 1550 nm:
Between 0 dBm and -55 dBm: ± 0.1 dB
< -55 dBm: ± 0.2 dB
3. Linearity for 850 dBm:
Between 0 dBm and -45 dBm: ± 0.2 dB
< -45 dBm: ± 0.25 dB

Table of Contents