DTX Series CableAnalyzer
Technical Reference Handbook
7-22
Running the Autotest in Smart Remote Mode
WCaution
If the reference test cords have been disconnected
from the tester’s or smart remote’s output since
the reference was set, you must set the reference
again to ensure valid measurements.
1
Turn on the tester and smart remote and let them sit
for 5 minutes. Allow additional time if the modules
have been stored above or below ambient
temperature.
2
Verify that the settings described in Table 7-2 on page
7-14 are appropriate.
3
Clean the connectors on the cabling to be tested.
4
Connect the tester to the cabling. Figure 7-10 shows
test connections for Method B.
5
Turn the rotary switch to AUTOTEST. Verify that the
media type is set to Fiber. Press
J Change Media to
change it if necessary.
6
Press P.
7
If Open or Unknown appears as the status, try the
following:
Verify that all connections are good.
Verify that the remote tester is on.
Try different connections to the cabling until the
test continues. See “Using FindFiber in Smart
Remote Mode” on page 7-44 for details on
FindFiber messages.
Use the visual fault locator to verify fiber continuity.
8
To save the results, press N, select or create a fiber ID
for one fiber; then press N.
Select or create a fiber ID for the other fiber; then press
N again.

Table of Contents

Other manuals for Fluke DTX-MFM2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke DTX-MFM2 and is the answer not in the manual?

Fluke DTX-MFM2 Specifications

General IconGeneral
BrandFluke
ModelDTX-MFM2
CategoryControl Unit
LanguageEnglish

Related product manuals