To run the Autotest on coaxial cabling:
A
Verify that the settings listed in Table 5-1 are
appropriate.
B
Attach coaxial adapters appropriate to the tester and
the smart remote.
C
Turn the rotary switch to AUTOTEST and turn on the
smart remote. Connect to the cabling, as shown in
Figure 5-3 for network cabling or Figure 5-4 for video
cabling.
If a fiber module is installed, you may need to press
J Change Media to select Coax as the media type.
D
Press Pon the tester or smart remote. To stop the test
at any time, press
I.
The tester shows the Autotest Summary screen when
the test is complete (see page 5-9). To view results for a
specific parameter, use A D to highlight the
parameter; then press H.
E
To save the results, press N. Select or create a cable ID;
then press
N again.
Note
If you turn off the main or remote unit while the
two units are connected through coaxial adapters,
the unit will turn on again.