EasyManua.ls Logo

Fluke DTX-MFM2 - Page 88

Fluke DTX-MFM2
320 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DTX Series CableAnalyzer
Technical Reference Handbook
3-10
Figure 3-5. Autotest Summary Screen for Twisted Pair Cabling
amd77f.eps
A
PASS: All parameters are within limits.
FAIL: One or more parameters exceeds the limit.
PASS*/FAIL*: One or more parameters are within the tester’s
accuracy uncertainty range, and the “*” notation is required by
the selected test standard. See “PASS*/FAIL* Results” on page 3-
12.
B
Press K or L to scroll the screen.
C
If the test failed, press J for diagnostic information.
D
Action prompt for the screen. Use D A to highlight a
parameter; then press H.
E
E: The measurement was within limits.
i: The parameter was measured, but has no PASS/FAIL limit in the
selected test limit. The results are for informational purposes only.
X: The measurement exceeds the limit.
U: See “PASS*/FAIL* Results” on page 3-12.
F
The worst margin found for the test.

Table of Contents

Other manuals for Fluke DTX-MFM2

Related product manuals