EasyManua.ls Logo

Fona XPan DG Plus - Checking Power Supply Adequacy

Fona XPan DG Plus
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FONA XPan DG Plus
47
1.12.4. Checking power supply adequacy
This test is done at the end of the installation and configuration to
determine the power supply adequacy during exposure.
It is done by measuring the line voltage drop during exposure, using
the parameters for a typical panoramic radiography.
Be sure the mains room switch and system mains switch are
turned OFF.
Select 300 VAC line voltage range on multi-meter and connect
measuring leads to the L and N points on the mains input block
XC501.
Turn ON the mains room switch and system mains switch.
Switch the unit ON.
Do Reset and wait for initialization.
Prepare the Pan Calibration Utility DG Suite program on PC and
establish exposure readiness.
Select P1 program
Select 73 kV / 10mA at control panel
RETURN Arm Movement to bring the arm position START,
ready to start the exposure
Run the acquisition program to receive the radiographic image
being created.
ATTENTION RADIATION Upon depression of the exposure key
radiation is generated.
Expose by keeping the exposure key pressed till end of movement.
During exposure read the minimum voltage under load.
After the end of the exposure read the voltage at no load.
Compute voltage drop under load (difference of voltage with no
load minimum voltage under load).
Max permissible voltage drop under load:
5 V at 230 VAC
5 V at 115 VAC
Turn unit OFF and remove meter leads.
Put back covers.
If the voltage drop is not within the specified, range advise the
customer, that an adequate power supply must be installed.
Disconnect unit and do not release for use.
1
Mains Fuse F501
2
Mains Input Block (starting from right Ground, Live and Neutral mains input)

Table of Contents

Related product manuals