2.2.7. Exposure setting
This function is intended to assign custom anode voltage (kV) and anode current (mA) or electrical charge (mAs)
values for all the available exposure programs.
By keeping pressed the BACK key for 5 s the system enters Service Mode.
Service function can be selected using the INCREASE or DECREASE keys on the
RIGHT side of the display.
Scroll the menu until Exposure Settings is displayed.
Press the PROGRAM key to enter Exposure Settings.
NOTE
Keep pressed the TEST key to reset all values to default (factory)
conditions.
Press the PROGRAM key to select the exposure program
for ADULT panorama (P1, P3, P4, P5, P6, P7)
for CHILD panorama (P2)
for CEPH (P8, P9, P10)
Select the patient build by pressing the PATIENT key,
Small, Medium, Large, Extra Large
PAN programs
Change the kV level to the desired value by pressing the INCREASE or
DECREASE keys on the left side of the display.
Change the mA level to the desired value by pressing the INCREASE or
DECREASE keys on the right side of the display.
Keep pressed the LIGHT key to store the selected value.
Repeat kV setting and mA setting for the other types of patient build if requested
for the other CHILD or ADULT group of programs if requested
Press the BACK key to exit Exposure Settings.
Press the TEST Mode key to exit SERVICE MODE.
Do verify proper setting by changing PATIENT build for ADULT
panorama (programs P1, P3, P4, P5, P6, P7), CHILD panorama (program
P2).