EasyManua.ls Logo

Fona XPan DG Plus - Axis

Fona XPan DG Plus
148 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Manuals
72
2.2.1.2. X Axis
NOTE
The X axis is a variable allowing to set the displacement offset of the rotating arm. The
plate is moved by stepped motor and controlled by three opto-sensors (encoder).
The right displacement of the rotating arm is reached when the canine laser falls on the
test phantom pin.
Use BACKWARD to move backward if the canine laser is after the test phantom pin.
Use FORWARD to move forward if the canine laser is behind the test phantom pin.
Press the PROGRAM key to enter PAN Alignment.
Possibility to operate on Rotation Axis is offered first
Press INCREASE or DECREASE key to change to X Axis
Press the PROGRAM key to enter setting Offset for X Axis
The actual offset value stored in memory is shown
Press RETURN key to enter X Axis edit mode
Press INCREASE or DECREASE keys on the LEFT side of the display
(large movements) or the two arrows (small movements) to shift the arm
backward or forward till the desired position.
Keep pressed the PROGRAM key to store new Offset X value
Press BACK key to exit X Axis
Press TEST Mode key to exit SERVICE MODE

Table of Contents

Related product manuals