Chapter 5. Error Codes and Troubleshooting
92 AquaTrans™ AT600 User’s Manual
5.1.2.5 E5: Amplitude Error
Problem: The signal amplitude exceeds the limits programmed in the Program > Advanced > Error Limits
> Amp Disc Min/Max
menu.
Cause: Solid or liquid particulates may be present in the flowcell. Also, poor coupling for the clamp-on
transducers can cause this problem.
Action: Refer to “Diagnostics” on page 93, to correct any flowcell problems.
5.1.2.6 E6: Cycle Skip
Problem: The fluid acceleration exceeds the limits programmed in the Program > Advanced > Error Limits
> Acceleration
menu.
Cause: This condition is usually caused by poor flow conditions or improper transducer alignment.
Action: Refer to “Diagnostics” on page 93, to correct any flowcell and/or transducer problems.