Chapter 4. Initial Setup and Programming
86 AquaTrans™ AT600 User’s Manual
4.7.10 Calculating the Path when Clamp-on selected
Select Path and press [√]. The following screen opens.
Press
[] again. Record this calculated transducer spacing for use in
installing your transducers on the pipe. The calculation is based on your
Sensor Setup input (pipe, transducer, fluid, and transducers).
Note: If setting the calculated spacing for your transducers is not
possible, contact the factory for assistance. Only if instructed by
the factory, press [
] again, and a screen similar to the following
opens.
Use the [] or [] key to select a specific digit, then use the [] or [] key
to change that digit as required. After all digits are correct, press
[] to
save. Then, press [
] to return to the previous screen.
IMPORTANT: Changing the transducer spacing should only be completed
according to instructions received from the factory.
Sensor Setup
Traverses
Fluid Type
Fluid Temperature
Path
Transducer Spacing
0.000 mm
Transducer Spacing
UNIT: mm
0.000
]MOVE
]UNDO [ ]SAVE
]
]
]MODF
[