AquaTrans™ AT600 User’s Manual 25
Chapter 3. Wiring the AT600 Electronics
3.2 Wiring the Line Power
The AT600 may be ordered for operation with power inputs of either 85-264 VAC or 12-28 VDC. The label on
the shroud inside the electronics enclosure lists the required line voltage for your meter.
Examples of AT600 labels indicating the required line voltage are shown in Figure 25 below.
Figure 25: Typical Meter S/N Label (AC and DC Versions)
ATTENTION EUROPEAN CUSTOMERS! To meet CE Marking requirements, all cables must be
installed as described in “Wiring Cable Specifications and Requirements” on page 177.
WARNING! Be sure to connect the AT600 only to the specified line voltage.
ATTENTION EUROPEAN CUSTOMERS! For compliance with the European Union's Low Voltage
Directive, this unit requires and external power disconnect device such as a switch or circuit breaker.
The disconnect device must be marked as such, clearly visible, directly accessible, and located within
1.8 m (6 ft) of the unit.