Chapter 6. Communication
140 AquaTrans™ AT600 User’s Manual
Command 197 (0xC5): Read Transducer Traverses and Spacing
This command is to read transducer traverses and spacing.
Table 110: Request Data Bytes for Read Transducer Traverses and Spacing
Byte Format Description
None
Table 111: Response Data Bytes for Read Transducer Traverses and Spacing
Byte Format Description
0 Unsigned-8 Transducer traverse
1 - 4 float Transducer spacing
Table 112: Command-Specific Response Codes for Read Transducer Traverses and Spacing
Code Class Description
0 Success No Command-Specific Errors
1-5 Undefined
6 Error Device-Specific Command Error
7-127 Undefined