AquaTrans™ AT600 User’s Manual 161
Chapter 6. Communication
Command 233 (0xE9): Write Signal Setup
This command is to write flow meter signal setup.
Table 170: Request Data Bytes for Write Signal Setup
Byte Format Description
0 Unsigned-8 signal setup type:
Delta T Offset
percentage Peak
Min Peak Percentage
Max Peak percentage
1 - 4 Float signal setup Value
Table 171: Response Data Bytes for Write Signal Setup
Byte Format Description
0 Unsigned-8 signal setup type:
Delta T Offset
percentage Peak
Min Peak Percentage
Max Peak percentage
1 - 4 Float signal setup Value
Table 172: Command-Specific Response Codes for Write Signal Setup
Code Class Description
0 Success No Command-Specific Errors
1Undefined
2 Error Invalid Selection
3-4 Undefined
5Error Too Few Data Bytes Received
6 Error Device-Specific Command Error
7Error In Write Protect Mode
8-15 Undefined
16 Error Access Restricted
17-127 Undefined