Revision C CASE 12-25
2060290-201
Exercise Test Setup
I Select the Max. Load Interpolation function (only possible when the exercise test
is performed on a bicycle ergometer). The maximum load is the load of the last
completed stage plus the duration of the last stage x load increase/stage time.
Example:
load of the last completed stage: 125 Watts
last stage aborted after 1 minute
load increase: 25 Watts
stage time: 2 minutes
125 W + 1 x 25/2 = 137.5 W
This value is also used to determine the percentage of the target load.
JSelect Save full disclosure ECG, then choose the phase when storage of the full
disclosure ECG should start, the sampling rate and whether or not you want to
record the audio ECG.
K Select the METS Interpolation Time.