12-56 CASE Revision C
2060290-201
System Configuration
Note
Use the GE Healthcare network installation kit to ensure all patient
isolation requirements are met.
A GE Healthcare Service Representative is required to configure the
MUSE system.
Only resting ECGs and exercise tests can be transferred to the
MUSE system.
The following resting ECG information is not transferred to the
MUSE system, because MUSE does not support this type of data:
hospital name, street, QT dispersion
arrhythmia labels for each QRS complex
vector loops
lead-related measuring values (Tabular Summary)
Full Disclosure ECG
I Select to transfer data to a MUSE system via a network, using a shared
directory (allows transfer of data to older MUSE versions via the network, a
separate shared directory is required for each workstation).
J Select to limit entered texts, such as in Patient Information, Test Information, to
the maximum length accepted by MUSE.
K Select to automatically transfer each test or each modification to a test to the
MUSE system.
L Select to automatically delete patient records from the local database after
transfer to a MUSE system.
M Select to start the modem connection before the transfer.
N Enter the MUSE Site number.
O Click to configure the Internet browser. Enter the path to the Internet browser
to start the MUSE Browser. Both the Internet browser (Internet Explorer,
version 4.0 and later) and the Acrobat Reader program must already be
installed. The Internet browser must be set up for proper communication (web
address of the MUSE system, etc.). Contact GE Healthcare Service for details.
P Enter location and cart/device number. Display the location name list.
Q Select the software version of the receiving MUSE system. Pre 5C, 5C to 7.1
or 7.2 and above. Versions V5C and above send a more detailed data record.
R Set the maximum permitted transmission time.
S Select to prevent adding of new texts in the Medication and Interpretation
windows. If deselected, texts can be added.
T Select to prevent adding of new texts in the Physicians and Technicians
windows. If deselected, texts can be added.
U Select to prevent editing of tests after transfer to the MUSE system.