D-8 CASE Revision C
2060290-201
Technical Description: Technical Specifications – CASE System
Handling of disposable supplies
and other consumables
Use only parts and accessories manufactured or recommended by GE
Healthcare.
Follow manufacturer’s instructions for use for disposable/consumable
product.
Follow local environmental guidelines concerning the disposal of
hazardous materials.
Patient mode of operation Continuous
Patient leakage current <10 µA
Degree of protection against
electrical shock
Type BF defibrillation protection for the patient cable (acquisition module)
Defibrillator recovery time The recovery time from defibrillator discharge is maximum 8 seconds.
Maintenance frequency
Recommended user daily visual inspection and cleaning.
Recommended six-month routine maintenance checks and test
procedures performed by qualified technical personnel.
Table 7. Safety (Continued)
Item Description